User’s Guide
Shop online at
omega.com
e-mail: [email protected]
For latest product manuals:
omegamanual.info
HH503
Digital Thermometer
INTRODUCTION
This instrument is a 4½ digit, compact-sized portable digital thermometer
designed to use external K-type and J-type thermocouples as temperature
sensor.TemperatureindicationfollowsReferenceTemperature/VoltageTables
(N.I.S.T. Monograph 175 Revised to ITS-90) for K-type and J-type thermo-
couples. One K-type thermocouple are supplied with the thermometer.
SAFETY INFORMATION
Itisrecommendedthatyoureadthesafetyandoperationinstructionsbefore
using the thermometer.
WARNING
To avoid electrical shock, do not use this instrument when working
voltages at the measurement surface over 24V AC or DC.
WARNING
To avoid damage or burns, do not make temperature measurement in
microwave ovens.
CAUTION
Repeated sharp flexing can break the thermocouple leads. To prolong
lead life, avoid sharp bends in the leads, especially near the connector.
The
symbol on the instrument indicates that the operator must refer to an
explanation in this manual.
1
SPECIFICATIONS
ELECTRICAL
Temperature Scale: Celsius or Fahrenheit user-selectable
Measurement Range:
J-TYPE -200°C to 1050°C, (-328°F to 1922°F)
K-TYPE -200°C to 1370°C, (-328°F to 2498°F)
Resolution: 0.1°C or 0.2°F
Accuracy: Accuracy is specified for operating temperatures over the range of
18°C to 28°C (64°F to 82°F), for 1 year, not including thermocouple error.
±(0.05% rdg + 0.3°C) -50°C to 1370°C
±(0.05% rdg + 0.7°C) -50°C to -200°C
±(0.05% rdg + 0.6°F) -58°F to 2498°F
±(0.05% rdg + 1.4°F) -58°F to -328°F
Temperature Coefficient: 0.1 times the applicable accuracy specification
per °C from 0°C to 18°C and 28°C to 50°C (32°F to 64°F and 82°F to 122°F).
Input Protection: 24V dc or 24V ac rms maximum input voltage on any
combination of input pins.
Maximum Differential Common Mode Voltage (Maximum Voltage be-
tween T1 and T2 during measurement): 1volt.
Reading Rate: one time per second.
Input Connector: Accepts SMP male miniature thermocouple connectors.
2
ENVIRONMENTAL
Ambient Operating Ranges: 0°C to 50°C (32°F to 122°F) <80% R.H.
Storage Temperature: -20°C to 60°C (-4°F to 140°F) <70% R.H.
GENERAL
Display:4½ digit liquid crystal display (LCD) with maximum reading of 19999.
Overload: "----.-" is display.
Battery: Standard 9V battery.
Battery Life: 100 hours typical.
Auto power off: The meter key switch inactive for more than 30 minutes,
press power key to resume operation.
Dimensions: 192mm(H) x 91mm(W) x 52.5mm(D)
Weight: 365g.
3
OPERATING INSTRUCTIONS
1
2
Power Switch
The key turns the thermometer on or off. In the data SET mode, can not
power off, must leave data SET mode then power off.
°C/°F Selecting the Temperature Scale
Readings are dual displayed in either degrees Celsius(°C) or degrees
Fahrenheit(°F). When the thermometer is turned on, it is set to the
temperature scale that was in use when the thermometer was last turned
off. To change the temperature scale, press the °C/°F key.
3
HOLD Mode (only Main display)
Pressing the HOLD key to enter the Data Hold mode, the "HOLD" annun-
ciator is displayed. When HOLD mode is selected, the thermometer held
the present readings and stops all further measurements.
Pressing the HOLD key again to cancel HOLD mode causing thermometer
to resume taking measurements.
In the MIN/MAX recording mode, press HOLD key to stop the recording.
Press HOLD key again to resume recording. (Previously recorded reading
are not erased).
4
4
5
Cold Junction Temperature Display mode (second display)
Press C.J. TEMP key to display the thermocouple input connector cold
junction temperature on second display.
Press C.J. TEMP key again to exit this mode.
K/J T1 Input Thermocouple Type Select
The K/J key switch the thermocouple input selects the K-type or J-type
thermocouple as input.
When the thermometer is turned on, it is set to the type selected that was
in use when the thermometer was last turned off.
6
MIN MAX with Time record Mode
Press MIN MAX key to enter the MIN MAX Recording mode, (displays the
Maximum reading with time, Minimum reading with time and Average
reading stored in record mode). In the this mode the automatic power-off
feature is disabled and
key, °C/°F key, REL key, SET key, Hi/Lo LIMITS
key, K/J, C.J. TEMP and all Recall keys are disabled.
The beeper emits a tone when a new minimum or maximum value is
recorded.
Present temperature reading displayed on second display.
Push MIN MAX key to cycle through the MAX, MIN and AVG readings.
If an overload is recorded, the averaging function is stopped and average
value display"----.-".
5
The true average of all the reading taken over at least 22 hours preiod can
be displayed. If 22 hours is exceeded, new averages are no longer
calculated. The last calcalated value is retained as the average reading,
but the actual minimum and maximum reading will continue to be
captured.
In the this mode, press HOLD key to stop the recording of readings, all
values are frozen, press again to restart recording.
To prevent accidental loss of MIN, MAX and AVG data, in this mode can
only be cancelled by pressing and hold down the MIN MAX key for 2
seconds to exit and erased recorded readings.
7
REL Relative mode
Pressing REL key to enter the Relative mode, zero the display, and store
the displayed Reading as a reference value and annunciator REL is
displayed. Present temperature reading displayed on second display.
Press REL key again to exit the relative mode.
The relative value can also be entered by the user. (See "SET mode" later
in this manual.)
When the desired Relative value has been entered, press REL key to
enter the Relative mode, press SET key use set Relative value as a
reference value. Press REL key again to exit the relative mode.
In the Relative mode, the value (can not >±3000.0 counts) shown on the
LCD is always the difference between the stored reference and the
present reading.
6
8
9
Sec. Minu. Selecting the Time scale
Reading the third displayed (Time) in either second or minute. When the
thermometer turned on, it is set to second. To change the Time scale, press
Sec. Minu. key.
Maximum Time reading is 100 hours. If 100 hours is exceeded, reset Time
to zero and restart.
SET mode (Relative value set, Time set and Hi/Lo Limits value set)
1. Press SET key to enter Relative valuse SET mode (Press ENTER key can
escape relative valuse set mode), REL set mode. = = = =.= is displayed in
main display.
Relative value is entered via overlay numbers, then press overlay ENTER
key, stored the relative value, enter Time set mode.
2. Time set mode, (Press ENTER key can escape Time set mode)
=.= = = : = = is displayed in second and third display. Time (hours, minutes,
seconds) value is entered via overlay numbers, then press overlay ENTER
key. Time start from set time value, enter Hi/Lo Limits value set mode.
3. Hi Limit value set mode,
is displayed (Press ENTER key can escape Hi
Limit value set mode), = = = =.= is displayed in main display, Hi Limit value
is entered via overlay numbers, then press overlay ENTER key, stored the
HiLimitvalue,enterLoLimitvaluesetmode(Press ENTERkeycanescape
Lo Limit value set mode). = = = =.= is displayed in main display, Lo Limit
value is entered via overlay numbers, then press overlay ENTER key,
stored the Lo Limit value and exit SET mode.
7
4. When the thermometer is turned on. The Relative set value and Hi/Lo Limits
set value that was in use when thermometer was last turned off set values.
10
Relative value Recall display mode.
PressRELRECALLkeytodisplaytheRelativesetvalueonseconddisplay.
Press REL RECALL key again to exit this mode.
11
Hi/Lo Limits value Recall display mode
PressHi/LoRECALLkeytodisplaytheHisetLimitvalueonseconddisplay.
Press Hi/Lo RECALL key again to display the Lo Limits set value on second
display. Press Hi/Lo RECALL key again to exit this mode.
12
Hi/Lo Limits mode (only Main display)
Press Hi/Lo LIMITS key to enter the Hi/Lo Limits comparative mode,
is
displayed, when input temperature value exceed Hi or Lo Limits value. The
beeper emits a continuity pulse tone. Press Hi/Lo Limits key again to exit
the Hi/Lo LIMITS mode.
8
OPERATOR MAINTENANCE
WARNING
To avoid possible electrical shock, disconnect the thermocouple
connectors from the thermometer before removing the cover.
Battery Replacement
Powerissuppliedbya9volt"transistor"battery. The" "appearsontheLCD
display when replacement is needed. To replace the battery, remove the two
screws from the back of the meter and lift off the battery cover. Remove the
battery from battery contacts.
9
CALIBRATION PROCEDURE
Note: The following calibration procedures should be performed only by
qualified technicians who have access to the following:
Equipment: The class of calibration should be 10 times greater than the
measured meter.
Step 1: Short the J3 with jumper.
Step 2: Apply 65 mV to the T1 input from Volts Calibrator and wait till the
reading becomes stable, then press the SET key until the display shows 0.1.
Step 3: Apply Type K 0°C to the input T2 and wait till the reading
becomes stable, then press the SET key until the meter shuts off by itself.
Step 4: Remove the jumper from J3.
10
NOTES:
11
NOTES:
12
WARRANTY/ DISCLAIMER
OMEGA ENGINEERING, INC. warrants this unit to be free of defects in materials and workmanship for a period of 13
months from date of purchase. OMEGA’s WARRANTY adds an additional one (1) month grace period to the normal
one (1) year product warranty to cover handling and shipping time. This ensures that OMEGA’s customers receive
maximum coverage on each product.
If the unit malfunctions, it must be returned to the factory for evaluation. OMEGA’s Customer Service Department will
issue an Authorized Return (AR) number immediately upon phone or written request. Upon examination by OMEGA, if
the unit is found to be defective, it will be repaired or replaced at no charge. OMEGA’s WARRANTY does not apply to
defects resulting from any action of the purchaser, including but not limited to mishandling, improper interfacing,
operation outside of design limits, improper repair, or unauthorized modification. This WARRANTY is VOID if the unit
shows evidence of having been tampered with or shows evidence of having been damaged as a result of excessive
corrosion; or current, heat, moisture or vibration; improper specification; misapplication; misuse or other operating
conditions outside of OMEGA’s control. Components in which wear is not warranted, include but are not limited to
contact points, fuses, and triacs.
OMEGA is pleased to offer suggestions on the use of its various products. However, OMEGA neither
assumes responsibility for any omissions or errors nor assumes liability for any damages that result from
the use of its products in accordance with information provided by OMEGA, either verbal or written.
OMEGA warrants only that the parts manufactured by the company will be as specified and free of defects.
OMEGA MAKES NO OTHER WARRANTIES OR REPRESENTATIONS OF ANY KIND WHATSOEVER,
EXPRESSED OR IMPLIED, EXCEPT THAT OF TITLE, AND ALL IMPLIED WARRANTIES INCLUDING ANY
WARRANTY OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE ARE HEREBY
DISCLAIMED. LIMITATION OF LIABILITY: The remedies of purchaser set forth herein are exclusive, and the
total liability of OMEGA with respect to this order, whether based on contract, warranty, negligence,
indemnification, strict liability or otherwise, shall not exceed the purchase price of the component upon
which liability is based. In no event shall OMEGA be liable for consequential, incidental or special damages.
CONDITIONS: Equipment sold by OMEGA is not intended to be used, nor shall it be used: (1) as a “Basic Component”
under 10 CFR 21 (NRC), used in or with any nuclear installation or activity; or (2) in medical applications or used on
humans. Should any Product(s) be used in or with any nuclear installation or activity, medical application, used on
humans, or misused in any way, OMEGA assumes no responsibility as set forth in our basic WARRANTY/ DISCLAIMER
language, and, additionally, purchaser will indemnify OMEGA and hold OMEGA harmless from any liability or damage
whatsoever arising out of the use of the Product(s) in such a manner.
RETURN REQUESTS/ INQUIRIES
Direct all warranty and repair requests/inquiries to the OMEGA Customer Service Department. BEFORE RETURNING
ANY PRODUCT(S) TO OMEGA, PURCHASER MUST OBTAIN AN AUTHORIZED RETURN (AR) NUMBER FROM
OMEGA’S CUSTOMER SERVICE DEPARTMENT (IN ORDER TO AVOID PROCESSING DELAYS). The assigned AR
number should then be marked on the outside of the return package and on any correspondence. The purchaser is
responsible for shipping charges, freight, insurance and proper packaging to prevent breakage in transit.
FOR NON-WARRANTY REPAIRS, consult OMEGA for
current repair charges. Have the following information
available BEFORE contacting OMEGA:
1. Purchase Order number to cover the COST
of the repair,
2. Model and serial number of the product, and
3. Repair instructions and/or specific problems
relative to the product.
FOR WARRANTY RETURNS, please have the following
information available BEFORE contacting OMEGA:
1. Purchase Order number under which the product was
PURCHASED,
2. Model and serial number of the product under warranty,
and
3. Repair instructions and/or specific problems relative to
the product.
OMEGA’s policy is to make running changes, not model changes, whenever an improvement is possible. This affords our
customers the latest in technology and engineering. OMEGA is a registered trademark of OMEGA ENGINEERING, INC.
© Copyright 2008 OMEGA ENGINEERING, INC. All rights reserved. This document may not be copied, photocopied, reproduced,
translated, or reduced to any electronic medium or machine-readable form, in whole or in part, without the prior written consent
of OMEGA ENGINEERING, INC.
Where Do I Find Everything I Need for
Process Measurement and Control?
OMEGA…Of Course!
Shop online at omega.comsm
TEMPERATURE
DATA ACQUISITION
Ⅺ
ߜ Ⅺ
ߜ Ⅺ
ߜ Thermocouple, RTD & Thermistor
Probes, Connectors, Panels & Assemblies
Wire: Thermocouple, RTD & Thermistor
Calibrators & Ice Point References
Recorders, Controllers & Process
Monitors
Data Acquisition & Engineering Software
Communications-Based Acquisition
Systems
Ⅺ
ߜ Ⅺ
ߜ Ⅺ
ߜ Ⅺ
ߜ Plug-in Cards for Apple, IBM &
Compatibles
Ⅺ
ߜ Ⅺ
ߜ Datalogging Systems
Ⅺ
ߜ Infrared Pyrometers
Recorders, Printers & Plotters
PRESSURE, STRAIN AND FORCE
HEATERS
Ⅺ
ߜ Ⅺ
ߜ Ⅺ
ߜ Ⅺ
ߜ Ⅺ
ߜ Ⅺ
ߜ Ⅺ
ߜ Ⅺ
ߜ Ⅺ
ߜ Transducers & Strain Gages
Load Cells & Pressure Gages
Displacement Transducers
Instrumentation & Accessories
Heating Cable
Cartridge & Strip Heaters
Immersion & Band Heaters
Flexible Heaters
Laboratory Heaters
FLOW/LEVEL
Ⅺ
ߜ Rotameters, Gas Mass Flowmeters &
Flow Computers
ENVIRONMENTAL
MONITORING AND CONTROL
Ⅺ
ߜ Ⅺ
ߜ Ⅺ
ߜ Air Velocity Indicators
Ⅺ
ߜ Ⅺ
ߜ Ⅺ
ߜ Ⅺ
ߜ Ⅺ
ߜ Metering & Control Instrumentation
Refractometers
Turbine/Paddlewheel Systems
Totalizers & Batch Controllers
Pumps & Tubing
Air, Soil & Water Monitors
Industrial Water & Wastewater
Treatment
pH, Conductivity & Dissolved Oxygen
Instruments
pH/CONDUCTIVITY
Ⅺ
ߜ Ⅺ
ߜ Ⅺ
ߜ pH Electrodes, Testers & Accessories
Benchtop/Laboratory Meters
Controllers, Calibrators, Simulators &
Pumps
Ⅺ
ߜ Ⅺ
ߜ Industrial pH & Conductivity Equipment
M2808A/1008
|