Omron Healthcare Blood Pressure Monitor BP710 User Manual

INSTRUCTION MANUAL  
Automatic  
Blood Pressure Monitor  
Model BP710  
 
INTRODUCTION  
Thank you for purchasing the OMRON® BP710 IntelliSense® Automatic Blood  
Pressure Monitor.  
Fill in for future reference.  
DATE PURCHASED:__________________  
SERIAL NUMBER:____________________  
• Staple your purchase receipt here  
• Register your product on-line at  
Your new digital blood pressure monitor uses the oscillometric method of blood  
pressure measurement. This means the monitor detects your blood’s movement  
through your brachial artery and converts the movements into a digital reading. An  
oscillometric monitor does not need a stethoscope so the monitor is simple to use.  
The BP710 comes with the following components:  
• Monitor • Standard Adult Arm Cuff  
• Instruction Manual printed in English and Spanish • Quick Start Guide  
The BP710 IntelliSense® Automatic Inflation Blood Pressure Monitor is intended for  
home use.  
SAVE THESE INSTRUCTIONS  
Download from Www.Somanuals.com3. All Manuals Search And Download.  
SAfETY INfORMATION  
To assure the correct use of the product, basic safety measures should always be  
followed including the warnings and cautions listed in this instruction manual.  
SAFETY SYMBOLS USED IN THIS INSTRUCTION MANUAL  
Indicates a potentially hazardous situation which, if not avoided,  
could result in death or serious injury.  
WARNING  
Indicates a potentially hazardous situation which, if not avoided,  
may result in minor or moderate injury to the user or patient or  
damage to the equipment or other property.  
CAUTION  
OPERATING THE DEVICE  
Contact your physician for specific information about your blood pressure.  
Self-diagnosis and treatment using measured results may be dangerous. Follow the  
instructions of your physician or licensed healthcare provider.  
The monitor is not intended to be a diagnostic device.  
If battery fluid should get in your eyes, immediately rinse with plenty of clean  
water. Contact a physician immediately.  
Read all of the information in the instruction manual and any other literature in the  
box before operating the unit.  
This device is intended for use in measuring blood pressure and pulse rate in the  
adult population. Do not use this device on infants or persons who cannot express  
their intentions.  
Read the Special Conditions section of the instruction manual if your systolic  
pressure is known to be more than 220 mmHg. Inflating to a higher pressure than  
necessary may result in bruising where the cuff is applied.  
Operate the device only as intended. Do not use the device for any other purpose.  
Dispose of the device, components and optional accessories according to applicable  
local regulations. Unlawful disposal may cause environmental pollution.  
Do not use a cellular phone near the device. It may result in an operational failure.  
 
4
SAfETY INfORMATION  
Use only Omron authorized parts and accessories. Parts and accessories not  
approved for use with the device may damage the unit.  
Use only 1.5 V “AAA” alkaline batteries with this device. Do not use other types of  
batteries. This may damage the unit.  
CARE AND MAINTENANCE  
Do not subject the monitor to strong shocks, such as dropping the unit on the floor.  
Do not submerge the device or any of the components in water.  
Store the device and the components in a clean, safe location.  
Changes or modification not approved by the manufacturer will void the user  
warranty. Do not disassemble or attempt to repair the unit or components.  
Download from Www.Somanuals.com5. All Manuals Search And Download.  
BEfORE TAKING A MEASUREMENT  
To ensure a reliable reading follow these recommendations:  
1. Avoid eating, drinking alcohol, smoking, exercising, and bathing for 30 minutes  
before taking a measurement. Rest for at least 15 minutes before taking the  
measurement.  
2. Stress raises blood pressure. Avoid taking measurements during stressful times.  
3. Measurements should be taken in a quiet place.  
4. Remove tight-fitting clothing from your left arm.  
5. Sit on a chair with your feet flat on the floor. Rest your left arm on table so that the  
cuff is at the same level as your heart.  
6. Remain still and do not talk during the measurement.  
7. Wait 2-3 minutes between measurements. The wait time allows the arteries to return  
to the condition prior to taking the blood pressure measurement. You may need to  
increase the wait time depending on your individual physiological characteristics.  
8. Keep a record of your blood pressure and pulse readings for your physician. A  
single measurement does not provide an accurate indication of your true blood  
pressure. You need to take and record several readings over a period of time. Try to  
measure your blood pressure at the same time each day for consistency.  
 
6
KNOW YOUR UNIT  
Main Unit:  
Display  
Battery compartment  
START/STOP button  
Memory (M) button  
Air jack  
Components:  
INSTRUCTION MANUAL  
Automatic  
Blood Pressure Monitor  
Air tube  
Air plug  
Model BP710  
INTRODUCTION  
®
Thank you for purchasing the OMRON BP710 Automatic Blood Pressure Monitor.  
Fill in for future reference.  
DATE PURCHASED: __________________  
SERIAL NUMBER: ____________________  
Quick Start  
Guide  
Instruction  
Manual  
Standard Adult Arm Cuff  
Download from Www.Somanuals.com7. All Manuals Search And Download.  
UNIT DISPLAY  
Systolic blood pressure  
Diastolic blood pressure  
Heartbeat symbol  
Pulse display  
Memory symbol  
(Displayed when viewing  
values stored in memory)  
Deflation symbol  
Irregular Heartbeat symbol  
Battery low indicator  
 
8
DISPLAY SYMBOLS  
IRREGULAR HEARTBEAT SYMBOL (  
)
When the monitor detects an irregular rhythm two or more times during the  
measurement, the Irregular Heartbeat Symbol ( ) will appear on the display with the  
measurement values.  
Normal Heartbeat  
An irregular heartbeat rhythm is defined as a  
rhythm that is more than 25% slower or 25%  
Pulse  
faster from the average rhythm detected while the  
monitor is measuring the systolic and diastolic  
blood pressure.  
Blood  
pressure  
Irregular Heartbeat  
If the irregular heartbeat symbol (  
)
Short Long  
displays with your measurement results,  
we recommend you consult your  
physician. Follow the directions of your  
doctor.  
Pulse  
Blood  
pressure  
HEARTBEAT SYMBOL/HYPERTENSION INDICATOR (  
TAKING A MEASUREMENT  
)
The Heartbeat Symbol ( ) flashes on the display during the measurement. The  
Heartbeat Symbol ( ) flashes at every heartbeat.  
When the measurement is complete, the Heartbeat Symbol ( ) flashes on the  
display with your blood pressure and pulse rate if the reading is above 135 for the  
Systolic Blood Pressure value and/or above 85 for the Diastolic Blood Pressure  
value.  
USING THE MEMORY FUNCTION  
The Heartbeat Symbol ( ) flashes when the blood pressure reading stored in the  
memory is above 135 for the Systolic Blood Pressure value and/or above 85 for the  
Diastolic Blood Pressure value.  
Download from Www.Somanuals.com9. All Manuals Search And Download.  
DISPLAY SYMBOLS  
The American Heart Association recommends the following guideline for upper limit  
of normal for home blood pressure readings:  
AHA Home Guideline for Upper Limit of Normal BP  
Systolic Blood Pressure  
Diastolic Blood Pressure  
135 mmHg  
85 mmHg  
This is a general guideline because blood pressure depends on age and morbidity.  
The AHA recommends a lower home blood pressure goal for certain patients,  
including diabetic patients, pregnant women, and patients with renal failure1.  
WARNING  
The monitor is not intended to be a diagnostic device.  
WARNING  
Contact your physician for specific information about your blood pressure.  
Self-diagnosis and treatment using measured results may be dangerous.  
Follow the instructions of your physician or licensed healthcare provider.  
Omron Healthcare recommends you contact your physician if the Heartbeat Symbol  
(
) flashes on the display with your measurement values.  
1American Heart Association Scientific Statement 2004  
 
10  
BATTERY INSTALLATION  
CAUTION  
Use only 1.5V “AAA” alkaline batteries with this device.  
1. Press the ▲ indicator on the battery cover and slide  
the cover off in the direction of the arrow.  
2. Install 4 “AAA” size batteries so the + (positive) and -  
(negative) polarities match the polarities of the battery  
compartment as indicated.  
NOTE: 4 “AAA” size batteries are not included in this  
unit.  
3. Replace the battery cover.  
BATTERY REPLACEMENT  
Low Battery Indicator  
When the Low Battery Indicator appears on the display, turn the monitor off and  
remove all the batteries. Replace with four new batteries at the same time. Long-life  
alkaline batteries are recommended.  
WARNING  
If battery fluid should get in your eyes, immediately rinse with plenty of  
clean water. Contact a physician immediately.  
CAUTION  
Dispose of the device, components and optional accessories according to applicable  
local regulations. Unlawful disposal may cause environmental pollution.  
Download from Www.Somanuals.co1m1. All Manuals Search And Download.  
APPLYING THE ARM CUff  
CAUTION  
Read all of the information in the instruction manual and any other literature in the  
box before operating the unit.  
CAUTION  
This device is intended for use in measuring blood pressure and pulse rate in the  
adult population. Do not use this device on infants or persons who cannot express  
their intentions.  
APPLYING THE CUFF ON THE LEFT ARM  
1. Make sure the air plug is securely inserted in the main unit.  
2. Remove tight-fitting clothing from your upper arm.  
3. Apply the cuff to your left upper arm so  
the arrow is centered on the inside of your  
arm and aligned with your middle finger.  
The air tube runs down the inside of your  
arm. The bottom of the cuff should be  
approximately 1/2" above your elbow.  
Left Arm  
1/2 inch (1 - 2 cm)  
Arrow  
Air Tube  
 
12  
APPLYING THE ARM CUff  
4. Wrap the cuff firmly in place around your  
arm using the cloth fastening.  
APPLYING THE CUFF ON THE RIGHT ARM  
When taking a measurement using the right arm, use this instruction for Step 3.  
Right Arm  
Apply the cuff to your right upper arm so the  
arrow and the air tube runs along the inside  
of your arm. The bottom of the cuff should be  
approximately 1/2" above your elbow.  
Arrow  
NOTE: Be careful not to rest your arm on the air tube. This will restrict the flow of air  
to the cuff.  
Download from Www.Somanuals.co1m3. All Manuals Search And Download.  
TAKING A MEASUREMENT  
1. Sit on a chair with your feet flat on the floor.  
Place your arm on a table so the cuff is level  
with your heart.  
Keep still and do not talk during measurement.  
2. Press the START/STOP button.  
All symbols appear on the display.  
The cuff starts to inflate automatically. As the cuff inflates, the monitor  
automatically determines your ideal inflation level.  
Remain still and do not move your arm until the entire measurement process is completed.  
NOTE: To stop the inflation or measurement, push the START/STOP button. The  
monitor will stop inflating, start deflating, and will turn off.  
3. Inflation stops automatically and the measurement is started.  
As the cuff deflates, decreasing numbers appear on the display and the Heartbeat  
Symbol ( ) flashes.  
4. When the measurement is complete, the arm cuff completely  
deflates. Your blood pressure and pulse rate are displayed.  
 
14  
TAKING A MEASUREMENT  
5. Press the START/STOP button to turn the monitor off.  
NOTE: The monitor will automatically turn off after five minutes.  
SPECIAL CONDITIONS  
CAUTION  
Inflating to a higher pressure than necessary may result in bruising where the cuff is  
applied.  
If your systolic pressure is known to be more than 220 mmHg, press and hold the  
START/STOP button until the monitor inflates 30 to 40 mmHg higher than your  
expected systolic pressure.  
NOTES: • Do not apply more pressure than necessary.  
• The monitor will not inflate above 299 mmHg.  
When the measurement is complete, the arm cuff completely deflates.  
Your blood pressure and pulse rate are displayed.  
Download from Www.Somanuals.co1m5. All Manuals Search And Download.  
USING THE MEMORY fUNCTION  
Every time you complete a measurement the monitor stores the blood pressure and pulse  
rate in the memory. The monitor automatically stores up to 14 sets of measurement values  
(systolic pressure, diastolic pressure, and pulse rate). When 14 sets of measurement  
values are stored, the oldest record is deleted to save the most recent values.  
1. Press the Memory (M) button.  
The most recent set of measurement values appears on the display.  
The values are displayed from the most recent to the oldest.  
NOTE: The number of the reading stored in the memory appears for a second  
before the pulse rate is displayed. The newest set is numbered “1”.  
2. Press the START/STOP button to turn the unit off.  
TO DELETE ALL THE VALUES STORED IN MEMORY  
You cannot partially delete values stored in the memory. All values will be deleted.  
1. Press the Memory (M) button.  
Then while holding it down, press the START/STOP button down for more than 2  
seconds until the display appears as illustrated below.  
2. Press the START/STOP button to turn the unit off.  
NOTE: The monitor will automatically turn off after five minutes.  
 
16  
CARE AND MAINTENANCE  
To keep your digital blood pressure monitor in the best condition and protect the unit  
from damage, follow the directions listed below:  
Do not forcefully bend the arm cuff or air tube. Do not fold tightly.  
Clean the monitor with a soft dry cloth.  
Do not use any abrasive or volatile cleaners.  
Do not attempt to clean the cuff.  
CAUTION  
Do not submerge the device or any of the components in water. Do  
not subject the monitor to extreme hot or cold temperatures,  
humidity or direct sunlight.  
CAUTION  
Store the device and the components in a clean, safe location.  
CAUTION  
Do not subject the monitor to strong shocks, such as dropping the unit on the floor.  
Remove the batteries if the unit will not be used for three months or longer. Always  
replace all the batteries with new ones at the same time.  
Use the unit consistent with the instruction provided in this manual.  
Use only authorized parts and accessories.  
Use the unit in a manner consistent with the instructions provided in this manual.  
CAUTION  
Use only Omron authorized parts and accessories. Parts and accessories not  
approved for use with the device may damage the unit.  
CAUTION  
Changes or modification not approved by the manufacturer will void the user  
warranty. Do not disassemble or attempt to repair the unit or components.  
Download from Www.Somanuals.co1m7. All Manuals Search And Download.  
ERROR INDICATORS  
SYMBOL  
CAUSE  
CORRECTION  
Take another measurement and  
remain still until the measurement  
is complete.  
Monitor could not  
detect pulse wave.  
Cuff under-inflated.  
Cuff not applied  
correctly.  
Remove the arm cuff.  
Read “Taking a Measurement”.  
Wait 2-3 minutes.  
Take another measurement.  
Cuff over-inflated.  
Batteries are worn.  
Device error.  
Replace the four batteries. Refer to  
“Battery Installation”.  
Contact Customer Service.  
 
18  
TROUBLESHOOTING TIPS  
PROBLEM  
CAUSES AND SOLUTIONS  
No power.  
Replace all four batteries with new ones.  
No display appears on the unit. Check the battery installation for proper placement  
of the battery polarities.  
Blood pressure varies constantly. Many factors  
including stress, time of day, how you wrap the  
cuff, may affect your blood pressure.  
Review the sections “Before Taking a Measure-  
Measurement values appear too  
high or too low.  
ment” and “Taking a Measurement”.  
Download from Www.Somanuals.co1m9. All Manuals Search And Download.  
fCC STATEMENT  
Note:  
POTENTIAL FOR RADIO/TELEVISION INTERFERENCE (for U.S.A. only)  
This product has been tested and found to comply with the limits for a Class B digital  
device, pursuant to part 15 of the FCC rules.  
These limits are designed to provide reasonable protection against harmful interference  
in a residential installation. The product generates, uses, and can radiate radio  
frequency energy and, if not installed and used in accordance with the instructions,  
may cause harmful interference to radio communications. However, there is no  
guarantee that interference will not occur in a particular installation. If the product does  
cause harmful interference to radio or television reception, which can be determined by  
turning the product on and off, the user is encouraged to try to correct the interference  
by one or more of the following measures:  
• Reorient or relocate the receiving antenna.  
• Increase the separation between the product and the receiver.  
• Connect the product into an outlet on a circuit different from that to which the  
receiver is connected.  
• Consult the dealer or an experienced radio/TV technician for help.  
POTENTIAL FOR RADIO/TELEVISION INTERFERENCE (for Canada only)  
This digital apparatus does not exceed the Class B limits for radio noise emissions  
from digital apparatus as set out in the interference-causing equipment standard entitled  
“Digital Apparatus”, ICES-003 of the Canadian Department of Communications.  
Cet appareil numérique respecte les limites de bruits radioeléctriques applicables aux  
appareils numériques de Clase B prescrites dans la norme sur le materiel brouilleur:  
“Appareils Numériques”, ICES-003 édictée par le minister des communications.  
Changes or modifications not expressly approved by the party responsible for  
compliance could void the user’s authority to operate the equipment.  
 
20  
LIMITED WARRANTIES  
Your BP710 IntelliSense® Automatic Blood Pressure Monitor, excluding the arm cuff,  
is warranted to be free from defects in materials and workmanship appearing within  
2 years from the date of purchase, when used in accordance with the instructions  
provided with the monitor. The arm cuff is warranted to be free from defects in  
materials and workmanship appearing within one year from the date of purchase when  
the monitor is used in accordance with the instructions provided with the monitor. The  
above warranties extend only to the original retail purchaser.  
We will, at our option, repair or replace without charge any monitor or arm cuff  
covered by the above warranties. Repair or replacement is our only responsibility and  
your only remedy under the above warranties.  
To obtain warranty service contact Customer Service by calling 1-800-634-4350 for the  
address of the repair location and the return shipping and handling fee.  
Enclose the Proof of Purchase. Include a letter, with your name, address, phone  
number, and description of the specific problem. Pack the product carefully to prevent  
damage in transit. Because of possible loss in transit, we recommend insuring the  
product with return receipt requested.  
THE FOREGOING IS THE SOLE WARRANTY PROVIDED BY OMRON IN  
CONNECTION WITH THIS PRODUCT, AND OMRON HEREBY DISCLAIMS  
ANY OTHER WARRANTIES, EXPRESS OR IMPLIED, INCLUDING  
IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A  
PARTICULAR PURPOSE. IMPLIED WARRANTIES AND OTHER TERMS  
THAT MAY BE IMPOSED BY LAW, IF ANY, ARE LIMITED IN DURATION  
TO THE PERIOD OF THE ABOVE EXPRESS WARRANTY.  
OMRON SHALL NOT BE LIABLE FOR LOSS OF USE OR ANY OTHER  
SPECIAL, INCIDENTAL, CONSEQUENTAIL OR INDIRECT COSTS,  
EXPENSES OR DAMAGES.  
This warranty provides you with specific legal rights, and you may have other rights  
that vary by jurisdiction. Because of special local requirements, some of the above  
limitations and exclusions may not apply to you.  
FOR CUSTOMER SERVICE  
Visit our web site at:  
Call toll free:  
1-800-634-4350  
Download from Www.Somanuals.co2m1. All Manuals Search And Download.  
SPECIfICATIONS  
BP710 (HEM-7113-Z)  
LCD Digital Display  
Model:  
Display:  
Pressure: 0 to 299 mmHg,  
Pulse: 40 to 180/min.  
Measurement Range:  
Pressure: ±3mmHg or 2% of reading  
Pulse: ±5% of reading  
Accuracy/Calibration:  
Automatic by electric pump  
Inflation:  
Automatic pressure release valve  
Automatic exhaust valve  
Deflation:  
Rapid Air Release:  
Measurement Method:  
Power Source:  
Battery Life:  
Oscillometric method  
1.5V 4 “AAA” batteries  
Approx. 300 uses with 4 new alkaline batteries  
50ºF to 104ºF (10ºC to 40ºC) /15 to 90% RH  
Operating Temperature /  
Humidity:  
-4ºF to 140ºF (-20ºC to 60ºC) /10 to 95% RH  
700-1060 hPa  
Storage Temperature /  
Humidity/Air Pressure:  
Approximately 8 5/8 oz. (245g)  
not including batteries  
Main Unit Weight:  
Main Unit Dimensions:  
Cuff Size:  
Approx. 5"(l) × 4 1/8"(w) × 2 1/2" (h)  
(128 mm × 104 mm × 64 mm)  
Approximately 5 3/4"(w) × 17 1/2"(l) (146 mm × 446 mm)  
Cuff tube 23 5/8" (600 mm)  
Fits arm circumferences 9" to 13"  
(220 mm to 320 mm)  
Cuff Circumference:  
 
22  
SPECIfICATIONS  
Up to 14 readings  
Memory:  
Contents:  
Main Unit, Standard Adult Arm Cuff, Instruction  
Manual, and Quick Start Guide  
Large Cuff 13" - 17" (320 mm - 420 mm)  
Optional Accessories:  
= Type B  
NOTE: These specifications are subject to change without notice.  
Download from Www.Somanuals.co2m3. All Manuals Search And Download.  
Omron Healthcare, Inc.  
1200 Lakeside Drive  
Bannockburn, IL 60015  
Copyright © 2010 Omron Healthcare, Inc.  
Made in China  
5325502-1B  
 

Milwaukee Sander 6065 User Manual
MSi Computer Hardware MS 6834B User Manual
Napa Essentials Battery Charger 85 2500 User Manual
NEC Projector LT81 User Manual
Nikon Camera Flash SB 700 User Manual
Nokia Portable Speaker MD 6 User Manual
Optoma Technology Projector TX773 User Manual
Oreck Vacuum Cleaner ZIP100 User Manual
Oregon Scientific Network Card AC DC Adapter User Manual
Oster Oven TSSTTVRB04 User Manual